Z-0692-2012: Varian High Energy Linear Accelerator, C-Series Clinac, Reference/FSCA Identifier: CP-06611; Models Numbe

Recall record Z-0692-2012 for Varian Medical Systems, Inc. Oncology Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0692-2012
Company in FDA record
Varian Medical Systems, Inc. Oncology Systems
FDA classification date
Jan 11, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

An event has been reported to Varian which entails excessive connector resistance, which caused the actual jaw positions to differ from the intended jaw positions without warning the operator.

Source facts

Action
Varian Medical Systems sent an Urgent Medical Device Correction - Urgent Field Safety Notice dated November 28, 2011 ***Expanded to include product #2 on December 16, 2011***, to all affected customers. ***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to establish a comprehensive quality assurance program before beginning treatments for the day, quality assurance tests must be performed. Additionally, any discrepancies in imaging field sizes vs planned field size should be brought to the attention of the medical physicist for investigation before treatment continues. Customers were asked to complete the attached Proof of Notification or Receipt Verification Card once they have read the document and return it to Varian Medical Systems. For question regarding this recall: USA and Canada 1-888-827-4265 Europe +41 41 749 8844
Initiated
2011-11-28
Root cause
Device Design
Terminated
2015-06-02
Distribution
Worldwide Distribution
Recall number
Z-0692-2012
Recall status
Terminated
CDRH record ID
106351
FDA classified
2012-01-11
Recall event number
60659

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗