Open original FDA record ↗Recall record
- FDA identifier
- Z-0692-2019
- Company in FDA record
- Agfa-Gevaert, N.V.
- FDA record creation date
- Jan 7, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
After an upgrade of the software of the Overhead Tube Crane, there were isolated cases in which the Overhead Tube Crane movement does not stop when the movement button is released. Instead of stopping while the movement button is released, the Tube Head Crane moves to the intended position.
Source facts
- Action
- An Urgent Field Safety Notice dated 12/4/18 was sent to customers. Actions undertaken by Agfa: Soon Agfa or its representative will visit your site and install the latest revision of the hardware board, which will prevent a movement without user interaction. Recommended actions to be taken by you: If you notice an Overhead Tube Crane movement without user interaction, notify your Agfa Service contact at once. Please distribute this information within your facility to all those who need to be aware of it. It is important to take the actions detailed in this Urgent Field Safety Notification and to acknowledge receipt of this notification.
- Initiated
- 2018-12-04
- Root cause
- Software design
- Terminated
- 2020-04-24
- Distribution
- US Nationwide Distribution in the states: IN, KS, MA, SC, and VA.
- Recall number
- Z-0692-2019
- Recall status
- Terminated
- CDRH record ID
- 169827
- FDA record created
- 2019-01-07
- Recall event number
- 81780