Z-0692-2019: DX-D 600 DIRECT RADIOGRAPHY SYSTEM Product Usage: The DX-D 600 system is a ceiling mounted General Radio

Recall record Z-0692-2019 for Agfa-Gevaert, N.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0692-2019
Company in FDA record
Agfa-Gevaert, N.V.
FDA record creation date
Jan 7, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

After an upgrade of the software of the Overhead Tube Crane, there were isolated cases in which the Overhead Tube Crane movement does not stop when the movement button is released. Instead of stopping while the movement button is released, the Tube Head Crane moves to the intended position.

Source facts

Action
An Urgent Field Safety Notice dated 12/4/18 was sent to customers. Actions undertaken by Agfa: Soon Agfa or its representative will visit your site and install the latest revision of the hardware board, which will prevent a movement without user interaction. Recommended actions to be taken by you: If you notice an Overhead Tube Crane movement without user interaction, notify your Agfa Service contact at once. Please distribute this information within your facility to all those who need to be aware of it. It is important to take the actions detailed in this Urgent Field Safety Notification and to acknowledge receipt of this notification.
Initiated
2018-12-04
Root cause
Software design
Terminated
2020-04-24
Distribution
US Nationwide Distribution in the states: IN, KS, MA, SC, and VA.
Recall number
Z-0692-2019
Recall status
Terminated
CDRH record ID
169827
FDA record created
2019-01-07
Recall event number
81780

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗