Skip to content

Public FDA record

Z-0693-2007: BioHorizons 0.050 Hex Driver, Regular, REF 300-350, BioHorizons Implant Systems, Inc., Birmingham, AL 35

Recall record Z-0693-2007 for BioHorizons Implant Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0693-2007
Company in FDA record
BioHorizons Implant Systems Inc
FDA classification date
Apr 13, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Premature Wear; The material from which the hex driver was made was improperly hardened during manufacturing and could lead to premature wear.

Source facts

Action
Letters and replacement drivers were sent to the consignees (both domestic and abroad) on 3/12/2007 explaining the problem and requesting that the consignee discard the instrument and use the replacement. They were delivered via United Parcel Service (UPS) 3rd day package delivery.
Initiated
2007-03-12
Root cause
Other
Terminated
2007-09-04
Distribution
Worldwide Distribution -- USA and countries of Canada, Norway, Australia, Spain, Brazil, Pakistan, Israel, Latvia, Japan, Columbia, Thailand, Turkey, Italy, South Korea, India, Taiwan, Ukraine, Russia, United Arab Emirates, Saudi Arabia, Hong Kong, Romania, UK, Kuwait, and Lithuania
Recall number
Z-0693-2007
Recall status
Terminated
CDRH record ID
51322
FDA classified
2007-04-13
Recall event number
37687

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗