The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause injury, delay in treatment, or Increased Intraperitoneal Volume (IIPV) (overfill).
Source facts
Action
Fresenius Medical Care issued on December 15, 2015, a n Important Product Notification to all affected customers informing them of the problem and how to recognize the proper latch closing. Questions regarding the contents of the Customer Notification or questions regarding the issue, contact Technical Services at 1-800-227-2572 and reference the Field Action Number FA-2015-10-W.
A reply form is included as part of the customer communication with
instructions to fill out and return via fax or scanned/emailed copy.
Initiated
2015-12-15
Root cause
Device Design
Terminated
2019-10-07
Distribution
Nationwide Distribution
Recall number
Z-0693-2016
Recall status
Terminated
CDRH record ID
142511
FDA classified
2016-01-27
Recall event number
72926