Skip to content

Public FDA record

Z-0693-2018: Coated VICRYL RAPIDETM (Polyglactin 910) Suture

Recall record Z-0693-2018 for Ethicon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0693-2018
Company in FDA record
Ethicon, Inc.
FDA record creation date
Feb 23, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

On August 17, 2017, Ethicon had decided to initiate a medical device recall (removal) of the twenty-five (25) lots of VICRYL RAPIDETM (Polyglactin 910) Suture that were distributed to the field. During finished goods testing, a suture damage was identified on a lot of VICRYL RAPIDETM Suture. During the course of the subsequent investigation, the suture also failed knot tensile strength testing.

Source facts

Action
On August 31, 2017 a Medical Device recall letter (Reference 2210968-08.31.17-003-R) was issued requesting customers to do the following: 1. Examine your inventory immediately to determine if you have product subject to this recall (removal) on hand and quarantine such product(s). 2. Remove the product subject to this recall (removal) and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall (removal) has been forwarded to another facility, contact that facility to arrange return. If you have additional questions regarding this recall (removal) or to report any customer complaints, please contact Ethicon Customer Support Center at 1-877-ETHICON (1-877-384-4266). As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA's MedWatch Adverse Event Reporting program online, by regular mail or by fax: " Online: www.fda.gov/medwatch/report.htm " Regular Mail: Use postage-paid FDA form 3500 available at: www.fda.gov/MedWatch/getforms.htm Mail to: MedWatch 5600 Fishers Lane, Rockville, MD 20852-9787 " Fax: 1-800-FDA-0178
Initiated
2017-08-17
Root cause
Equipment maintenance
Terminated
2019-05-08
Distribution
Nationally
Recall number
Z-0693-2018
Recall status
Terminated
CDRH record ID
161312
FDA record created
2018-02-23
Recall event number
79086

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗