Reason stated in the source
Breathing circuit reservoir bag may be assembled at the wrong attachment point, which will result in less than optimal therapy.
Source facts
- Action
- On 3/28/23, correction notices were emailed to customers who were asked to remove and switch the reservoir bag to the proper location. On 1/24/24, follow-up correction notices were sent to customers who were asked to do the following: 1) Review your inventory, and if affected devices are identified, ensure the reservoir bag is in the proper location. 2) If affected devices were further distributed, send a copy of this notice to your customers. 3) Complete and return the Effectiveness Check Questionnaire via email to customerservice.us@sentec.com If you have any questions, contact the firm's Customer Service at customerservice.us@sentec.com or by phone at (208)- 263- 2549 Monday - Friday from 7 AM to 3:30 PM (PST). Additional information for contact information can be found at www.sentec.com/contact
- Initiated
- 2023-03-28
- Root cause
- Process control
- Distribution
- US: RI, CA, TX
- Recall number
- Z-0693-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205048
- FDA classified
- 2024-01-17
- Recall event number
- 93648