Open original FDA record ↗Recall record
- FDA identifier
- Z-0693-2025
- Company in FDA record
- Noah Medical
- FDA classification date
- Dec 18, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
Source facts
- Action
- On 10/25/24, the firm started hand delivering correction notices to customers informing them of the following: If your device is affected the firm will contact you to inspect and if necessary to replace the power supply. Firm contact: customerservice@noahmed.com, 1-888-325-6624
- Initiated
- 2024-10-25
- Root cause
- Device Design
- Distribution
- US: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NY
- Recall number
- Z-0693-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210988
- FDA classified
- 2024-12-18
- Recall event number
- 95712