Skip to content

Public FDA record

Z-0694-2013: Spacelabs Smart Disclosure System, Model 92810, is an integrated component of the Intesys Clinical Suite

Recall record Z-0694-2013 for Spacelabs Healthcare, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0694-2013
Company in FDA record
Spacelabs Healthcare, Llc
FDA classification date
Jan 17, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Displaying or printing of an incorrect waveform for the time indicated in the printout or display in the Spacelabs Healthcare Intesys Clinical Suite (ICS) G2, Version 4.03, Clinical Access portion of Smart Disclosure, Model 92810.

Source facts

Action
Spacelabs has sent the Urgent -Medical Device Correction Spacelabs Healthcare ICS G2 Version 4.03 Model 92810 Smart Closure letter, dated 19 December 2012, to all U.S. customers. The firm plans to send a customer letter (translated as necessary) via e-mail to all international subsidiaries and distributors of record on December 27, 2012. The letter is addressed to Hospital Administrator/Biomedical Manager/Office Manager. Users are advised on the following actions: 1) Advise your staff of this potential failure and advise them to always click the "Refresh" button when scrolling to new waveform data prior to printing or viewing waveform information or storing saved events. 2) We will contact you at the earliest possible date to schedule a convenient time for your hospital to upgrade your affected software at no cost. Customers with questions can contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.
Initiated
2012-12-05
Root cause
Software design
Terminated
2013-07-08
Distribution
Worldwide Distribution-USA including Puerto Rico and the states of California, Florida, Iowa, Illinois, Louisiana, Michigan, Minnesota, North Carolina, South Carolina, Texas, and Washington, and the countries of Argentina, Mexico, France, and India.
Recall number
Z-0694-2013
Recall status
Terminated
CDRH record ID
115067
FDA classified
2013-01-17
Recall event number
63931

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗