Z-0694-2019: Ingenia Elition S, Ingenia Elition X

Recall record Z-0694-2019 for Philips Medical Systems Nederlands. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0694-2019
Company in FDA record
Philips Medical Systems Nederlands
FDA record creation date
Jan 8, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During automatic tabletop movement, pressing and holding the Stop Table button may lead to an unexpected restart of the table movement.

Source facts

Action
On December 4, 2018, the firm notified customers of the recall via an Urgent Medical Device Correction letter. The letter informed customers of the product issue. Customers were asked to do the following: The user must follow the Instructions for Use to verify that no part of the patient's body, hair, clothing, cables or infusion lines can get trapped or injured by any part of the equipment. Additionally, the VitalScreen shows positioning guidance regarding arm supports and padding. The user should avoid the use of the yellow Stop Table button on the VitalScreen. The user must always use the tumble switch or physical Emergency Table Stop button for cases which require the table movement to stop. Philips stated they would be rolling out FCO78100490 Service Pack 1 for R5.5.0 to affected systems. The Service Pack will remove the "Stop Table" software button on the VitalScreen.
Initiated
2018-12-03
Root cause
Under Investigation by firm
Terminated
2020-05-12
Distribution
CA, MN, NJ, NY, OH, TN, TX. Foreign distribution to Canada, Belgium, France, Germany, INdia, Japna, Netherlands, South Africa, Sweden, Switzerland, UAE.
Recall number
Z-0694-2019
Recall status
Terminated
CDRH record ID
169900
FDA record created
2019-01-08
Recall event number
81797

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗