Z-0694-2021: Revolution CT ES - Product Usage: intended for head, whole body, cardiac and vascular X-ray Computed Tomo

Recall record Z-0694-2021 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0694-2021
Company in FDA record
GE Healthcare, LLC
FDA record creation date
Dec 22, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A potential for protocol(s) with unintended patient orientation or scan setting when using Intelligent Protocoling

Source facts

Action
A Recall notification letter titled, "URGENT MEDICAL DEVICE CORRECTION" was sent to consignees on 11/23/2020 via a parcel service. The letter instructs the consignee that they can continue to use their Revolution CT ES or Revolution CT with Apex Edition. They are asked to ensure that users review all protocols recommended by the system. It is important to confirm that the proper protocol is selected, and the patient orientation properly matches the protocol orientation picture in the Scan Settings window. If using the Intelligent Protocoling Autoselect preference, and more than one protocol is shown in the Selected Protocols window, the users are asked to carefully review these protocols to only select the appropriate one. The firm mentions that GE Healthcare will correct all affected products at no cost to the consignee. A GE Healthcare representative will contact the consignee to arrange for the correction. If the consignee has any questions, they are to contact GE Healthcare Service at 1-800-437-1171 or their local Service Representative.
Initiated
2020-11-23
Root cause
Software in the Use Environment
Terminated
2023-08-29
Distribution
Worldwide distribution - US Nationwide distribution including in the states of FL and TX. The countries of Belgium, France, and Spain.
Recall number
Z-0694-2021
Recall status
Terminated
CDRH record ID
184853
FDA record created
2020-12-22
Recall event number
86932

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗