Skip to content

Public FDA record

Z-0694-2024: SoClean 3, REF SC1400, Automated Supplemental Sleep Equipment Maintenance System

Recall record Z-0694-2024 for SoClean, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0694-2024
Company in FDA record
SoClean, Inc
FDA classification date
Jan 10, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

New User Manual with additional instructions for use and a hose and mask adapter provided by the device manufacturer. This voluntary recall is intended to help reduce potential health risks of exposure to ozone gas after the use of a SoClean device.

Source facts

Action
An Urgent Medical Device Correction notice was issued November 16, 2023 advising customers: 1) Do not use your SoClean2 or SoClean3 without the hose and mask adapter provided by the manufacturer. 2) Review revised labeling and User Manuals for: additional warnings and set-up instructions to enable appropriate use; additional instructions for events in which a user is unable to smell residual ozone; additional clarity and consistency regarding that the SoClean2 and SoClean3 are not intended to replace CPAP manufacturers cleaning instructions but rather are to be used to supplement cleaning procedures for home use CPAP masks and tubing. 3) With each SoClean filter purchase, SoClean is supplying a complementary (no additional cost) Hose and Mask Adapter, which facilitates use of the SoClean2 and SoClean3 equipment without ozone entering the CPAP. Set-up Instructions for the Hose and Mask Adapter are included in the updated manuals. To obtain a no cost Hose and Mask Adapter, please contact our Customer Care team at 866-501-3705 or visit the SoClean website: https://www.soclean.com/field-correction. 4) As stated in the SoClean User Manual, patients with underlying lung diseases, such as asthma and chronic obstructive pulmonary disease (COPD, including emphysema and chronic bronchitis), and those with cardiovascular disease, may be sensitive to ozone and should consult with their health care provider before using a SoClean device. Customers are asked to confirm receipt of the recall notification online at the through the hyperlink embedded in the notice https://www.soclean.com/field-correction. Contact Customer Care at 866-501-3705 with questions.
Initiated
2023-11-15
Root cause
Labeling design
Distribution
Worldwide distribution - United States Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, Colombia, Egypt, El Salvador, England, France, Germany, Hong Kong, Isle of Man, Israel, Italy, Japan, Jordan, Korea, Lithuania, Mexico, New Zealand, Portugal, Russia, Saudi Arabia, Seoul, Singapore, Spain, Sri Lanka, Switzerland, Thailand, Netherlands, Turkey, United Arab Emirates, and United Kingdom.
Recall number
Z-0694-2024
Recall status
Open, Classified
CDRH record ID
204869
FDA classified
2024-01-10
Recall event number
93570

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗