Under certain conditions, * an excessive amount of anticoagulant may be infused into the patient. (* interruption of a self-test by an alarm and obstruction of the access lines)
Source facts
Action
Customers were notified by letter on 12/19/2006. They were told that Gambro would schedule the required upgrades to their equipment and not to use the current syringe for delivering anticoagulant.
Initiated
2007-01-04
Root cause
Device Design
Terminated
2012-01-03
Distribution
Nationwide distribution -- including states of AL, CA, FL, GA, IN, MD, ME, MI, MS, NC, NH, NY, OH, PA, SC, TX, VA, and WI.
Recall number
Z-0695-2007
Recall status
Terminated
CDRH record ID
50323
FDA classified
2007-03-31
Recall event number
37306