Z-0695-2020: Artis zee/ zeego Interventional Fluoroscopic X-Ray System, Model Number: 10094142, Software Version: VC21

Recall record Z-0695-2020 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0695-2020
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Dec 17, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In affected Artis zee floor Interventional Fluoroscopic X-Ray Systems, the activation of a collision sensor will cause a block the movement of the floating tabletop and could result in a delay or interruption of a procedure.

Source facts

Action
On November 14, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail stating: URGENT: MEDICAL DEVICE CORRECTION Dear Customer, This letter is to inform you of a corrective action that will be performed to prevent a possible hazard to patients, operators, or other persons and equipment due to a potential software error in software version VC21C and serial numbers 140422, 140423 or 140430. Artis zee systems with software version VC21C and serial # 140422, 140423 or 140430 This potential software issue has no influence on the previous treatment/diagnosis of patients. However, in case of software failure, planned procedures may have to be terminated and perfomed on an alternative x-ray system. What is the underlying issue requiring corrective action and when does the issue occur? In affected Artis systems, the movement of the floating tabletop may become blocked after a collision sensor has been activated during system movements. The collision supervision will display a user message on the monitor. What is the impact on system operation and what are the potential risks? During normal operation, activation of a collision sensor will cause a block of any system movements but moving the floating tabletop is still possible. In affected Artis systems, the movement of the floating tabletop is blocked and could result in a delay or interruption of a procedure. We do not consider it necessary to re-examine any patients in this case as this possible software error had no influence on treatment of patients. How was the issue detected, what is the cause, and what action will be taken? The issue was detected during regular field monitoring. The root cause for the blocked tabletop movement (horizontally and vertically) is a software error of the stand control unit. By using the safety override function and moving out of the collision zone, the system movement can be resumed. In safety override mode you may try to resolve a coll
Initiated
2019-11-14
Root cause
Software design
Terminated
2020-05-12
Distribution
US: WA and TX International: Denmark
Recall number
Z-0695-2020
Recall status
Terminated
CDRH record ID
178168
FDA record created
2019-12-17
Recall event number
84370

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗