Skip to content

Public FDA record

Z-0695-2022: Medtronic TurboHawk Plus Directional Atherectomy System 6F, Small Vessel, Catalog number THP-S. For use

Recall record Z-0695-2022 for ev3 Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0695-2022
Company in FDA record
ev3 Inc.
FDA classification date
Mar 7, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The device has similarities in design to another device that the firm recently issued a field correction on due to instances of tip detachment and embolization.

Source facts

Action
The recalling firm began issuing the URGENT MEDICAL DEVICE NOTICE dated February 2022 on 2/4/2022 via 2-day UPS delivery. The letter explained the issue and instructed the consignee to adhere to the existing IFU instructions and warnings and precautions listed in the letter to reduce the risk of guidewire prolapse. The Customer Instructions included the notice was to be shared with all who needed to be aware within the consignee's organization or to any other organization where the product has been transferred; to review the IFU included with the product prior to use, noting the warnings and precautions listed in the notice; and to complete the Customer Confirmation Form to acknowledge receipt of the notice and that the appropriate action has been taken. It was to be emailed to the recalling firm.
Initiated
2022-02-04
Root cause
Labeling design
Terminated
2024-10-02
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IA, IL, KS, LA, MI, MS, NC, NJ, NY, OR, SC, SD, TN, TX, and VA.
Recall number
Z-0695-2022
Recall status
Terminated
CDRH record ID
191705
FDA classified
2022-03-07
Recall event number
89470

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗