Z-0696-06: Fisher & Paykel Healthcare Adult Respiratory Ventilator Circuit-Dual Heated Catalogue Number RT210

Recall record Z-0696-06 for Fisher & Paykel Healthcare Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0696-06
Company in FDA record
Fisher & Paykel Healthcare Inc
FDA classification date
Mar 29, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Risk of fire. The manufacturing process may have the potential to damage the insulation of the heater wire in circuits within the specified lot range. This may lead to a potential risk of fire.

Source facts

Action
On December 20th, 2005 the firm sent the notification letter instructing consignees to check and destroy any stock in the affected lot range. A fax-back Device Recall Response Form was included with the letter and will be used for the company''s effectiveness checks.
Initiated
2005-12-20
Root cause
Other
Terminated
2012-04-19
Distribution
Nationwide in AR, CA, FL, IL, LA, MN, NC, OH, SD, TX, UT, & VA. International distribution
Recall number
Z-0696-06
Recall status
Terminated
CDRH record ID
43335
FDA classified
2006-03-29
Recall event number
34248

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗