Open original FDA record ↗Recall record
- FDA identifier
- Z-0696-2013
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Jan 18, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
A software security issue with Oracles TNS-Listener component has been identified. TNS-Listener is supplied with the Roche COBAS INTEGRA 400/400 plus Analyzers. The Oracle TNS-Listener can be considered as an authentication and redirection component between the client and the database inside the Roche product. This means an attacker could register their own non-Roche database via remote access
Source facts
- Action
- Roche sent an Urgent Medical Device Correction letter on November 19, 2012, to all affected customers via UPS Ground (receipt required). The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete the attached form and return by fax and file this Medical Device Correction for future reference. For technical support customers were instructed to contact Diagnostics Technical Support at 1-900-428-2336. For questions regarding this recall call 317-521-3911.
- Initiated
- 2012-11-19
- Root cause
- Software design
- Terminated
- 2014-01-13
- Distribution
- Nationwide Distribution including AL, AR, AZ, CA, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY and GUAYNABO, PR.
- Recall number
- Z-0696-2013
- Recall status
- Terminated
- CDRH record ID
- 114368
- FDA classified
- 2013-01-18
- Recall event number
- 63617