Z-0696-2020: smith&nephew PROFIX Mallet, 750 GRAMS, REF 71512452 - Product Usage: The PROFIX Mallet is a reusable surg

Recall record Z-0696-2020 for Smith & Nephew, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0696-2020
Company in FDA record
Smith & Nephew, Inc.
FDA record creation date
Dec 17, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a a potential failure mode associated with the use of the mallet, wherein the poly tip of the mallet may become loose and/or come off during use. If the potential failure mode occurs, particles generated in the poly thread potentially could enter the surgical field or a delay in surgery could occur if the tip disengages during the use of the mallet.

Source facts

Action
The firm disseminated a medical device field correction notice on 11/11/2019 by letter. The letter explained the failure and requested the following actions be taken by hospital representatives: 1. Please check your inventory and complete the attached Inventory Correction Certification Form. 2. If you have the PROFIX 750 Gram Mallet, please maintain awareness of this notice. 3. When preparing for cleaning, disassemble the poly tip from the mallet and inspect disassembled components for damage or wear. 4. After cleaning, inspect disassembled components and ensure the components can be re-assembled securely. 5. Ensure the PROFIX 750 Gram Mallet is sterilized in the disassembled state. 6. After sterilization, ensure the poly tip is re-assembled to the mallet securely.
Initiated
2019-11-11
Root cause
Device Design
Terminated
2021-05-07
Distribution
Worldwide distribution - US Nationwide distribution and countries of Belgium, Canada, China, Columbia, Costa Rica, Finland, France, Germany, Great Britain, Hong Kong, Italy, Netherlands, Norway, Panama, Poland, Switzerland, United Arab Emirates, Venezuela.
Recall number
Z-0696-2020
Recall status
Terminated
CDRH record ID
178075
FDA record created
2019-12-17
Recall event number
84325

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗