Open original FDA record ↗Recall record
- FDA identifier
- Z-0697-2016
- Company in FDA record
- Phadia Ab
- FDA classification date
- Jan 28, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
All equivocal and positive results (greater or equal 7 EliA U/ml) on EliA SmDP well lot 0018 may be incorrect and must be considered invalid. Patient samples in complaint investigations caused unspecific signals up to 22 U/ml on EliA SmDP well lot 0018. The nonspecific signals are not caused by anti-Sm antibodies, nor by streptavidin antibodies.
Source facts
- Action
- On 12/28/2015, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions call 1-800-3464364, option #2.
- Initiated
- 2015-12-28
- Root cause
- Nonconforming Material/Component
- Terminated
- 2016-08-16
- Distribution
- US Nationwide in the states of NC, GA, NJ, NY, PA, TX, IL, MI, MT, NM, and OR.
- Recall number
- Z-0697-2016
- Recall status
- Terminated
- CDRH record ID
- 142606
- FDA classified
- 2016-01-28
- Recall event number
- 72969