Z-0697-2019: ddRCombi Plus FP, Model Number 8.2000.0070.0 (US Version), 8.2000.0060.0 (Europe Version) The ddRCombi P

Recall record Z-0697-2019 for Swissray Medical AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0697-2019
Company in FDA record
Swissray Medical AG
FDA classification date
Jan 8, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A component of the device was recalled by a supplier. In rare cases of inadequate maintenance or very high clinical workload, the first steel wire rope of the lifting column, which is designed to take the load, can break without triggering the safety lock. This can lead to overloading of the safety cable, which is not designed to permanently carry the load of the lifting column or continuously move under load. This could cause material fatigue and cause the arm to drop unexpectedly during patient positioning. If the arm drops unexpectedly, patients and users may be seriously injured.

Source facts

Action
A component of the device was recalled by a supplier (Siemens AG/Siemens Healthcare GmbH), as a result, Swissray Medical AG, sent an "Important Warning / Important Safety Notice" letters dated 11/14/18 to its customers. The letter described the product, problem and actions to be taken. The firm stated safety related retrofits are available for installation to eliminate risk to patients and users.The retrofit aims to completely correct the problem. The customers were instructed to do the following: To mitigate risk to patients and users until the retrofit modification is implemented, users are strongly advised not to move the column up/down directly above the patient, but to instead first perform vertical movement next to the patient, and then horizontally position over the patient at a fixed height. Please ensure that all your users who own the affected products and others who are aware or them receive the safety-related information provided with this safety notice, and comply with the recommendations contained therein. If you have any questions, contact US Agent of Swissray Medical AG, QA/RA Coordinator at 908-372-6408 or email: Deniz.Kortan@swissray.com.
Initiated
2018-11-14
Root cause
Component design/selection
Terminated
2021-09-01
Distribution
Worldwide distribution: US distribution to states of: IL and MA and countries of: Czech Republic, Germany, Poland and Peru.
Recall number
Z-0697-2019
Recall status
Terminated
CDRH record ID
169713
FDA classified
2019-01-08
Recall event number
81749

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗