Z-0697-2020: Low Volume Cartridge IFU (SKU CAR-125-B) used with the following devices: 1) NxStage One, Model NX1000-1

Recall record Z-0697-2020 for NxStage Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0697-2020
Company in FDA record
NxStage Medical, Inc.
FDA record creation date
Dec 18, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential patient health risk while performing chronic hemodialysis treatments in the home setting with for patients with low blood volume (2.4 Liters or less in a patient with body weight of 30kg or less) when prescribing larger dialysate volumes.

Source facts

Action
On August 27, 2019, NxStage issued URGENT MEDICAL DEVICE CORRECTION notices to customers. Actions to be Taken by Customers: For patients using NxStage System One Cycler, model NX1000-1: - At any dialysate flow rate, the system may have a fluid removal error defined as the greater of 3% of Total Dialysate Volume or 300 ml/12hr. The Total Allowable Error is generally dependent on total dialysate volume processed for this System model. For patients using less than 10L of dialysate per treatment, this model has a minimum Total Allowable Error volume of 300 ml. For patients using NxStage System One S or NxStage VersiHD Cycler, model NX1000-3, NX1000-3A, NX1000-10, NX1000-10A, NX1000-16, or NX1000-16A: - At dialysate flow rates of 3.0 Liters/hour and below, the system may have a fluid removal error defined as the greater of up to 25 ml/hour or 5% of Total Ultrafiltration Volume. The Total Allowable Error is dependent on the error rate and length of treatment and can continue to increase for as long as treatment is being delivered. - At dialysate flow rates of above 3.0 Liters/hour, the system may have a fluid removal error defined as the greater of up to 100 ml/hour or 5% of Total Ultrafiltration Volume. The Total Allowable Error is dependent on the error rate and length of treatment and can continue to increase for as long as treatment is being delivered. NxStage discourages the use of the System for the treatment of low blood volume patients. If the physician considers it appropriate to prescribe home hemodialysis using the NxStage System One for low blood volume patients, the following points should be considered: 1. When performing hemodialysis treatments for low blood volume (2.4 Liters or less in a patient with body weight of 30kg or less) patients, NxStage strongly recommends employment of additional monitoring vigilance. Always ensure that cardiac status and fluid status are monitored for low blood volume patients by monitoring bloo
Initiated
2019-08-27
Root cause
Under Investigation by firm
Terminated
2021-02-01
Distribution
AL, AZ, CA, NC & VA
Recall number
Z-0697-2020
Recall status
Terminated
CDRH record ID
176185
FDA record created
2019-12-18
Recall event number
83706

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗