Z-0697-2021: Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) and FS 2-15 (Ref no. 615H515) in combination

Recall record Z-0697-2021 for Haag-Streit USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0697-2021
Company in FDA record
Haag-Streit USA Inc
FDA record creation date
Dec 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software error -Software REF 615 588 versions 2.0 to 3.3, movement of the focusing is triggered by pressing the button of the footswitch it might occur that the movement does not stop when releasing the button, result in potential risk to the patient's eye

Source facts

Action
Haag-Streit USA, Inc. notified accounts by Urgent Medical Device Correction letter via UPS on 11/20 and delivered 11/24, stating reason for recall, health risk, and action to take: Since the error only occurs when using the optional microscope accessory EIBOS 2, we recommend the following action: 1. Operate the focusing only as described in the instructions for use 668 600 part 5.4 when the fundus observation module EIBOS 2 is swiveled in (see following extract from the above-mentioned Instructions for Use). 5.4 During Surgery Surgery starts at the anterior segment of the eye. The operating microscope is focused on the cornea or the iris. The EIBOS 2 is folded upwards. " Fold the EIBOS 2 down to observe the posterior segment of the eye. " Focus the EIBOS 2 exclusively via its operating elements for focusing. Do not use the focusing of the operating microscope. " Use the foot switch to position the EIBOS 2 by means of the XV-coupling over the patient's eye. " Fold the EIBOS 2 up to observe the anterior segment of the eye 2. If the error occurs, press the focus button on the footswitch again. By pressing once more, the software error is corrected and the focusing works normally again. It is not possible for the error to occur twice in a row. Attention: After 10 seconds of focusing, during which the microscope moves in the direction of the patient's eye, the EIBOS 2 spring suspension to protect the eye is at the end position and the EIBOS 2 is pressed into the eye. 3. Perform a software update of the floor stand. You will receive the software version 3.4 through your local Haag-Streit USA representative who will fix the error. 4. Please send us the enclosed Acknowledgement Letter filled in as soon as the software update is complete. Transmission of the information Please ensure that all users of the products mentioned are aware of and receive this urgent safety information. If you have passed the products to third parties, please forward them a c
Initiated
2020-11-20
Root cause
Software design
Distribution
US Nationwide distribution in the state of OK.
Recall number
Z-0697-2021
Recall status
Open, Classified
CDRH record ID
184345
FDA record created
2020-12-23
Recall event number
86707

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗