Open original FDA record ↗Recall record
- FDA identifier
- Z-0698-2019
- Company in FDA record
- Ion Beam Applications S.A.
- FDA record creation date
- Jan 8, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a risk of detachment of the Compact Gantry Rolling Floor.
Source facts
- Action
- On December 12, 2018, the firm notified its customers of the recall via an Urgent Medical Device Correction letter. Customers were informed of the product issue. Customers were reminded that nobody should stand on the rolling floor when the gantry is rotating, as specified in the Clinical User Guide. The recalling firm stated they would take the following actions: Short-term action: IBA will replace the rod end bearings attaching the cGRF to the nozzle via a preventive maintenance on all ProteusONE sites. Long-term action: A new design of the anchoring system of the cGRF to the nozzle will be deployed on all impacted sites via an upgrade kit by end of October 2019. Customers with questions should contact Sylviane BERGER Vigilance@iba-group.com +32 10 203 787, or Helpdesk +32 2 507 20 81 (available 24/7)
- Initiated
- 2018-12-12
- Root cause
- Device Design
- Terminated
- 2020-04-15
- Distribution
- Worldwide Distribution - US Nationwide in the states of LA and MI. Foreign distribution to France, Japan, and UK.
- Recall number
- Z-0698-2019
- Recall status
- Terminated
- CDRH record ID
- 169970
- FDA record created
- 2019-01-08
- Recall event number
- 81807