Z-0700-2017: Model 37751 Recharger Product Usage: The Medtronic 37751 Recharger is a hand-held device used to charge

Recall record Z-0700-2017 for Medtronic Neuromodulation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0700-2017
Company in FDA record
Medtronic Neuromodulation
FDA record creation date
Dec 3, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medtronic has identified an increased number of complaints from customers involving reports of Rechargers (Medtronic Model 37751 Recharger, which is included in Models 37754 and 97754 Charging Systems,Spinal Cord Stimulation)that are in an unresponsive error state, where the Recharger is non-functional with a blank display screen and is beeping every 5 seconds. Medtronic has determined all Rechargers manufactured starting in November 2014 (indicated by serial numbers beginning with NKA4 or NKU4) are more susceptible to this error state.

Source facts

Action
Consignees were sent a series of documents which included an introductory Medtronic and "Urgent Medical Device Correction" letters dated September 2016. The urgent letters included one for Deep Brain Stimulation (DBS) and one for Spinal Cord Stimulation(SCS) . Each letter was addressed to Healthcare Professional. Only physicians associated with both SCS and DBS therapy received the cover letter. The letters provided information on "Background, Issues Mitigation, Recommendation, Additional Information and Patient Notification (DBS). Requested consignees to complete and return the Physician /HCP Reply Form. For questions contact Medtronic Technical Services at 1-800-707-0933 weekdays 7 am to 6pm, CT.
Initiated
2016-10-03
Root cause
Software change control
Terminated
2022-05-04
Distribution
Worldwide Distribution - US Nationwide including DC OUS: Canada and others.
Recall number
Z-0700-2017
Recall status
Terminated
CDRH record ID
150430
FDA record created
2016-12-03
Recall event number
75458

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗