Open original FDA record ↗Recall record
- FDA identifier
- Z-0700-2017
- Company in FDA record
- Medtronic Neuromodulation
- FDA record creation date
- Dec 3, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medtronic has identified an increased number of complaints from customers involving reports of Rechargers (Medtronic Model 37751 Recharger, which is included in Models 37754 and 97754 Charging Systems,Spinal Cord Stimulation)that are in an unresponsive error state, where the Recharger is non-functional with a blank display screen and is beeping every 5 seconds. Medtronic has determined all Rechargers manufactured starting in November 2014 (indicated by serial numbers beginning with NKA4 or NKU4) are more susceptible to this error state.
Source facts
- Action
- Consignees were sent a series of documents which included an introductory Medtronic and "Urgent Medical Device Correction" letters dated September 2016. The urgent letters included one for Deep Brain Stimulation (DBS) and one for Spinal Cord Stimulation(SCS) . Each letter was addressed to Healthcare Professional. Only physicians associated with both SCS and DBS therapy received the cover letter. The letters provided information on "Background, Issues Mitigation, Recommendation, Additional Information and Patient Notification (DBS). Requested consignees to complete and return the Physician /HCP Reply Form. For questions contact Medtronic Technical Services at 1-800-707-0933 weekdays 7 am to 6pm, CT.
- Initiated
- 2016-10-03
- Root cause
- Software change control
- Terminated
- 2022-05-04
- Distribution
- Worldwide Distribution - US Nationwide including DC OUS: Canada and others.
- Recall number
- Z-0700-2017
- Recall status
- Terminated
- CDRH record ID
- 150430
- FDA record created
- 2016-12-03
- Recall event number
- 75458