Skip to content

Public FDA record

Z-0701-2007: SCANLAN International Mobin-Uddin Vein Holder, Sterile, Disposable, single-use, catalog #1001-761, Lot AZ

Recall record Z-0701-2007 for Scanlan International Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0701-2007
Company in FDA record
Scanlan International Inc
FDA classification date
Apr 3, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Incorrectly assembled by manufacturer

Source facts

Action
The consignees were notified of the product recall by letter dated February 19, 2007. Consignees were asked to return product to the manufacturer and replacement product is being offered.
Initiated
2007-02-19
Root cause
Other
Terminated
2007-06-27
Distribution
Worldwide distribution --- USA including states of IN, MO and country of Japan
Recall number
Z-0701-2007
Recall status
Terminated
CDRH record ID
50950
FDA classified
2007-04-03
Recall event number
37596

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗