Open original FDA record ↗Recall record
- FDA identifier
- Z-0701-2017
- Company in FDA record
- 3M Company - Health Care Business
- FDA classification date
- Dec 3, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
3M is recalling Universal Electrosurgical Pads because a report of an unintended material (process liner) in the product. This may prevent or impede the safe return of electrosurgical current following Electrosurgical Unit (ESU) activation. This defect has the potential to increase the risk of a patient burn directly under the pad or at an alternate site of the body.
Source facts
- Action
- 3M sent an Urgent Medical Device Recall letter dated October 26, 2016 to affected customers. The letter identified the affected product, problem, and actions to be taken. Customers were instructed to complete and return the attached Customer Product Return and Recall Form. Customers with questions can contact 3M Health Care Customer Helpline at 1-800-228-3957, 8:30 am - 7:00 pm EST.
- Initiated
- 2016-10-26
- Root cause
- Equipment maintenance
- Terminated
- 2017-08-01
- Distribution
- Worldwide Distribution - US Nationwide in the states of AR, AZ, CA, FL, GA, ID, IL, KY, NC, NY, TN, TX, VA and the countries of Canada, Chile, Costa Rica, Germany, Japan, Mexico and South Africa.
- Recall number
- Z-0701-2017
- Recall status
- Terminated
- CDRH record ID
- 150979
- FDA classified
- 2016-12-03
- Recall event number
- 75644