Open original FDA record ↗Recall record
- FDA identifier
- Z-0701-2024
- Company in FDA record
- Icotec Ag
- FDA classification date
- Jan 10, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.
Source facts
- Action
- icotec ag issued an URGENT Medical Device Recall notice to its consignees beginning on 11/28/2023 via email. The notice explained the problem with the device, the risk to health, and informed the medical facility that the affected product had been removed prior to the notice. Decision on whether patient-level follow-up is required should be made by the physician. Replacements will be provided where applicable. The firm expanded this recall on 12/27/2023 (notices delivered by US mail) to include additional lots and one additional product. Contact for questions: Laura Husken, laura.huesken@icotec.ch, Tel.: +41 78 438 78 73
- Initiated
- 2023-11-28
- Root cause
- Other
- Distribution
- US: AZ. MI, OH, CA, CO, TX, NY, NC, DE, PA
- Recall number
- Z-0701-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205441
- FDA classified
- 2024-01-10
- Recall event number
- 93577