The software program written for use in the automated microtiter plate reader (Bio-Tek ELx800) for use with the Pro-Trac II Tacrolimus ELISA kit was found to be faulty. Assay software program did not include the Calibrator Zero OD specification (STDO>=1.500) as specified in the product insert.
Source facts
Action
Customers were sent notification letters and effectiveness check form. recommended users manually veify results using product insert specifications.
Initiated
2004-04-12
Root cause
Other
Terminated
2006-07-22
Distribution
CA, Washington DC, FL, KY & Canada
Recall number
Z-0702-05
Recall status
Terminated
CDRH record ID
36821
FDA classified
2005-04-13
Recall event number
30929