Reason stated in the source
The double wrapping of this custom kit basin set is not supported by the current sterility assurance validation.
Source facts
- Action
- As only one customer is potentially impacted, in the interest of prompt action and notification, a teleconference was held on Friday, November 16, 2018 with the customer to communicate the impending actions. During this teleconference, the customer was informed that an HHE and field corrective action was in-process and that a study was being initiated to determine that their requested packaging configuration was acceptable. The customer also was informed that they should not continue to use the impacted double wrapped basin, and the components contained within, until the study was completed. Cardinal Health is following up on this teleconference with a written notification.
- Initiated
- 2018-11-16
- Root cause
- Unknown/Undetermined by firm
- Terminated
- 2019-11-04
- Distribution
- US state of Florida
- Recall number
- Z-0702-2019
- Recall status
- Terminated
- CDRH record ID
- 169742
- FDA record created
- 2019-01-09
- Recall event number
- 81756