Skip to content

Public FDA record

Z-0702-2022: da Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11

Recall record Z-0702-2022 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0702-2022
Company in FDA record
Intuitive Surgical, Inc.
FDA record creation date
Mar 1, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Conductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional surgery may be required to remove the metal tip. If the procedure proceeds without the metal tip, there is potential for energy leakage, which could cause tissue damage. If the metal tip remains in the patient, adhesions may form, which may lead to bowel obstruction and chronic abdominal pain.

Source facts

Action
On 12/23/21 recall notices were mailed to customers who were asked to do the following: 1. Inform necessary hospital personnel about this issue. 2. Locate all affected devices at your site and return to the recalling firm. 3. Inform affected personnel when the requested actions have been completed. 4. Complete and return the attached Acknowledgement Form 5. Please retain a copy of this letter and the acknowledgement form for your files. If you need further information or support contact your Clinical Sales Representative or contact Intuitive Customer Service: North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customerservice@intusurg.com Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or EUCS@instusurg.com South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) or support.korea@intusurg.com Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST) or csjapan@intusurg.com Taiwan: 886-2-27008181 (7:30am to 5:30pm CT) or CS.Taiwan@intusurg.com Distributors were asked to provide their customers with the information found in the recall notice; the notice the recalling firm sent to their customers was included as an attachment. Distributors are to collect the affected cannula reducers and return them to the recalling firm. Distributors were asked to complete and return a recall response form. They were also asked to follow-up with their customers who don't respond to the recall notice at least three times.
Initiated
2021-12-23
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - US Nationwide distribution in the states of California, Wisconsin, Virginia, Ohio, Texas, New York, Nevada, Connecticut, New Jersey, Massachusetts, Georgia, Florida, Indiana, North Carolina, Missouri, Pennsylvania, Michigan, Oregon, Kentucky, Hawaii, Tennessee, Alabama, Nebraska, Maryland, Oklahoma, Colorado, Arkansas, Arizona, Kansas, Illinois, South Carolina, Minnesota, North Dakota, Iowa, Louisiana, Washington, Delaware, Mississippi, New Mexico, District of Columbia, Idaho, New Hampshire, Rhode Island, West Virginia, South Dakota, Alaska, Maine, Utah, Montana, Vermont, Wyoming, Puerto Rico and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Luxembourg, Martinique, Monaco, Netherlands, Norway, Poland, Qatar, Reunion, Romania, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.
Recall number
Z-0702-2022
Recall status
Open, Classified
CDRH record ID
190985
FDA record created
2022-03-01
Recall event number
89274

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗