Open original FDA record ↗Recall record
- FDA identifier
- Z-0703-2018
- Company in FDA record
- Oakworks Inc
- FDA classification date
- Feb 23, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The foot control used on the OAKWORKS Inc., CFPM300, CFPM301, CFPM400, CFPM401, CFPMB301, CFUR301, CFUR401 and CFLU401 Imaging Tables can become lodged under the base or under the column of the table.
Source facts
- Action
- OAKWORKS, Inc. issued a MEDICAL DEVICE ADVISORY NOTICE dated 8/18/2017 to customers providing more information in regards to the Medical Device Advisory Notice sent on May 12, 2017. The advisory notice was initiated in response to reported issues with the foot control. OakWorks, Inc. customer service will place the order for the appropriate Foot Control Kit listed below based on what table is impacted and the date of manufacture. PN 2015-204 - Foot Control Kit - Standard Riser 2015-209 - Foot Control Kit - Thinner Riser 2015-210 - Foot Control Kit - Holder Only 2015-211 - Foot Control Kit - Riser Removal Customers with questions may contact 1-800-558-8850 or 1-717-235-6807.
- Initiated
- 2017-07-31
- Root cause
- Device Design
- Terminated
- 2019-03-29
- Distribution
- USA (nationwide) Distribution
- Recall number
- Z-0703-2018
- Recall status
- Terminated
- CDRH record ID
- 161387
- FDA classified
- 2018-02-23
- Recall event number
- 79107