A potential hazardous situation may occur if the collimator locking handle is not properly placed in the lock position and the warning circuit does not detect that the collimator is unlocked because of the patient proximity to the collimator during a clinical scan.
Source facts
Action
The firm initiated the medical device correction by letter on 10/0/2017. The letter instructed the consignee to ensure the collimator locking handles are in the locked position at all times.
GE Healthcare will inspect and if necessary correct all affected products.
Initiated
2017-10-30
Root cause
Other
Distribution
Worldwide
Recall number
Z-0703-2019
Recall status
Completed
CDRH record ID
169652
FDA record created
2019-01-09
Recall event number
81736