It the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to transfer from the SkyPlate detector to the system. The image remains in the memory of the detector, but cannot be transferred wirelessly or by use of the backup cable. To continue, the operator can reset the SkyPlate detector by removing its batteries, but the acquired image is lost and a
re-take is necessary.
Source facts
Action
Philips sent an Urgent- Field Safety Notice date December 11, 2017.
Philips plans to release a new HP firmware to the installed base. In addition the existing Access Point Hardware in the field needs to be patched. A Philips Service Engineer will contact the affected Customers when the Field Action Kit is available to be implemented. For further questions, please call (978) 659-7275.
Initiated
2017-12-11
Root cause
Under Investigation by firm
Terminated
2020-05-29
Distribution
Worldwide Distribution - USA (nationwide) Distribution, Internationally to Canada.
Recall number
Z-0704-2018
Recall status
Terminated
CDRH record ID
161432
FDA record created
2018-02-23
Recall event number
79139