Z-0705-2012: GE LOGIQ E9 Diagnostic Ultrasound Systems running software versions R3.1.0 and R3.1.1. The device is int

Recall record Z-0705-2012 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0705-2012
Company in FDA record
GE Healthcare, LLC
FDA classification date
Jan 11, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

GE Healthcare has recently become aware of a potential safety issue associated with the data of your LOGIQ E9 system that may impact patient safety. 1 - User may notice an image from a different patient, or an image from the current patient with title information from a different patient. 2 Incorrect measurements/calculations on images, when viewed on the off-line review station. As a re

Source facts

Action
The firm, GE Healthcare" sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated December 20, 2011 to its customers. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, Contact Information and actions to be taken. The customers were instructed to ensure that all potential users in their facility are made aware of this safety notification and recommended actions and to follow the safety instructions noted in the letter. If you have any questions or concerns, contact your local GEHC service representative at 1-800-321-7937 (USA/Canada).
Initiated
2012-01-03
Root cause
Software design
Terminated
2013-01-10
Distribution
Worldwide distribution: USA (nationwide) including states of: AK, AZ, ALR, CA, CO, CT, FL, GA, IL, IN, IA, LA, LME, MD, MA, MI, MN, MS, MO, NJ, NM, NY, NC, ND, OH, OK, OR, PA, PR, SC, SD, TIN, TX, UT, VA, WA, WV, WI, and WY; and countries of: AUSTRALIA, AUSTRIA, BELGIUM CANADA, CHILE, CHINA, CROATIA, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, NETHERLANDS, NEW ZEALAND, NORWAY, PANAMA, PHILIPPINES, PORTUGAL, REPUBLIC OF KOREA, ROMANIA, RUSSIAN FEDERATION,SAUDI ARABIA, SINGAPORE, SLOVAKIA, SPAIN,SWEDEN, SWITZERLAND, TAIWAN, and UNITED KINGDOM..
Recall number
Z-0705-2012
Recall status
Terminated
CDRH record ID
106445
FDA classified
2012-01-11
Recall event number
60773

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗