In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.
Source facts
Action
The firm notified their consignees about the issue on 01/04/2019 and will install additional fixation.
Initiated
2019-01-04
Root cause
Under Investigation by firm
Distribution
TX and WI, Germany, Denmark, Netherlands, United Kingdom
Recall number
Z-0706-2019
Recall status
Completed
CDRH record ID
170222
FDA record created
2019-01-09
Recall event number
81886