Z-0706-2022: Everest MI XT Outer Dilator, Catalog Number 5101-90168

Recall record Z-0706-2022 for K2M, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0706-2022
Company in FDA record
K2M, Inc
FDA record creation date
Mar 1, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Stryker received two (2) complaints for units from specific lots of Inner and Outer Dilators not fitting together properly during a surgical procedure. Some dilators from these lots and from other lots were subsequently determined to be affected by a manufacturing nonconformance.

Source facts

Action
URGENT: MEDICAL DEVICE RECALL notification letters dated 1/13/22 were sent to customers. Hospital Quality or Materials Managers/Inventory Contacts instructions: Actions needed 1. Immediately check your internal inventory to locate the product listed on the attached Business Reply Form and remove them from their point of use. 2. Use the Business Reply Form to reconcile any affected product. Complete the Business Reply Form even if there is no affected product identified. 3. Return the enclosed Business Reply Form by email to Spine-RegulatoryActions@Stryker.com or fax to 201- 575-4675 to confirm receipt of this notification/document product segregation. Your account s Stryker Spine Sales Representative will arrange for retrieval and removal of any instruments associated with this recall. 4. In the interim, until units can be removed from your facility by Stryker, there are no additional actions that users can and should take once the product has been segregated and removed from point of use. - If you desire additional training associated with these instructions, please contact our Regulatory Compliance team by calling 201.749.8090 or by email at Spine-RegulatoryActions@Stryker.com, or contact your local Sales representative. 5. Maintain awareness of this communication internally until all required actions have been completed within your facility. 6. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients appropriately. 7. Under 21 CFR 803, manufacturers are required to report any serious injuries where a product has contributed or may have contributed to the event. Please keep Stryker informed of any adverse events associated with this product by using Stryker s online reporting site: https://www.stryker.com/productexperience. Additionally, adverse reactions or quality problems experienced with the use of this product may also be reported to the FDA's MedWatch Adver
Initiated
2022-01-13
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide.
Recall number
Z-0706-2022
Recall status
Open, Classified
CDRH record ID
191849
FDA record created
2022-03-01
Recall event number
89522

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗