Open original FDA record ↗Recall record
- FDA identifier
- Z-0707-2013
- Company in FDA record
- DePuy Spine, Inc.
- FDA classification date
- Jan 24, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head
Source facts
- Action
- DePuy Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated November 7, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory, quarantine the material and return remaining inventory. Customers were advised to complete the enclosed Business Reply Form and fax back to 508-828-3762 or email to lregan3@its.jnj.com. For questions call Linda Regan @508-977-6606.
- Initiated
- 2012-11-07
- Root cause
- Component change control
- Terminated
- 2013-12-24
- Distribution
- Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.
- Recall number
- Z-0707-2013
- Recall status
- Terminated
- CDRH record ID
- 114495
- FDA classified
- 2013-01-24
- Recall event number
- 63665