Skip to content

Public FDA record

Z-0707-2019: Siemens Biograph Horizon PET/CT System, Product Usage: The Siemens Biograph systems are combined X-Ray

Recall record Z-0707-2019 for Siemens Medical Solutions USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0707-2019
Company in FDA record
Siemens Medical Solutions USA, Inc.
FDA record creation date
Jan 10, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Improper design or specifications.

Source facts

Action
Siemens sent a letter to all their customers. The letter identified the affected product, problem and actions to be taken. Siemens Healthineers' service organization will contact you inQ1 of 2019 to schedule the replacement of the 70 amp CT circuit breaker, found within the PDCC cabinet of your Biograph Horizon system, with a 60 amp CT circuit breaker. Please ensure that this customer advisory notice is placed in the Biograph Horizon Operator's Manual for Examination and Acquisition and disseminated to all operators of Biograph Horizon. If this equipment is no longer in your possession, we kindly ask that you forward this letter to the new owner of the equipment and please inform Siemens Healthineers about the change in ownership. If you have any questions regarding this advisory notice, please contact your local Siemens Healthineers representative at the contact numbers provided below. " America: 1-800-888-7436 " Europe, Middle East, and Africa: +49 9131 940 4000 " Asia and Australia: +86 (21) 3811 2121
Initiated
2018-11-30
Root cause
Device Design
Terminated
2020-09-02
Distribution
Worldwide Distribution - US Nationwide in the states of AZ, CA, CO, DE, FL, GA, ID, IL, IN, KY, LA, MD, ME, MN, NJ, NM, NV, NY, OH, OK, PA, TN, TX, and VA and the countries of Argentina, Australia, Bahrain, Belarus, Belgium, Brazil, Chile, China, Colombia, France, Germany, India, Iran, Ireland, Italy, Japan, Lebanon, Mexico, Netherlands, New Zealand, Philippines, South Africa, Spain, Sri Lanka, Taiwan, Turkey, and United Kingdom
Recall number
Z-0707-2019
Recall status
Terminated
CDRH record ID
170001
FDA record created
2019-01-10
Recall event number
81821

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗