Colonoscope insertion tube is not bending properly and is stiff.
Source facts
Action
Fujinon sent out a Technical Bulletin on 3/27/2006 that instructs the physician to do a bending test with the colonoscope. If the scope does not bend around an 8 inch bending diameter, the scope is not to be used.
Initiated
2006-03-27
Root cause
Other
Terminated
2007-12-03
Distribution
Product is distributed nationwide to hospitals and physicians. It was shipped internationally to Canada. US Government accounts include VA medical centers in Spokane WA, St. Louis, MO and Camp Pendleton Naval Hospital in Camp Pendleton, CA.
Recall number
Z-0708-06
Recall status
Terminated
CDRH record ID
44461
FDA classified
2006-04-05
Recall event number
34680