Z-0708-2024: CORI REAL INTELLIGENCE Robotic Drill, Product No. ROB10013 and ROB10013S

Recall record Z-0708-2024 for Blue Belt Technologies, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0708-2024
Company in FDA record
Blue Belt Technologies, Inc
FDA classification date
Jan 11, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Possibility that the user is unable to resolve drill disconnection error messages.

Source facts

Action
The recalling firm sent an Urgent Medical Device Recall/Correction notice, dated 11/13/23, by mail to consignees. The notice informed consignees that they may continue to use affected devices as is provided that they ensure the backup Robotics Instrument Tray is available during surgery and to review the CORI Surgical System User Manual, which provides instructions for solving error messages. If consignees are unable to resolve the error messages, they should be prepared to convert to manual surgical procedure following the instructions in Appendix D: Recovery Procedure Guidelines of the CORI Surgical System User Manual. In accordance with routine servicing, consignees should return the CORI drill after 75 uses for servicing, during which the recalling firm will make the hardware update required to correct this recall event. Consignees are asked to contact their Account Representative to report pervasive drill disconnection error messages. Completed acknowledgement forms are to be returned to the recalling firm by email at fieldactions@Smith-Nephew.com. Questions are to be directed by email to fieldactions@smith-nephew.com or to your Account Representative.
Initiated
2023-11-13
Root cause
Device Design
Distribution
Domestic: Nationwide Distribution.
Recall number
Z-0708-2024
Recall status
Open, Classified
CDRH record ID
204976
FDA classified
2024-01-11
Recall event number
93620

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗