Z-0708-2025: Monitoring Service Application (MSA). Software to process, analyze, display, and report cardiac events in

Recall record Z-0708-2025 for Braemar Manufacturing, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0708-2025
Company in FDA record
Braemar Manufacturing, LLC
FDA classification date
Dec 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Not all Electrocardiogram (ECG) events received July 2022-July 2024 were not properly routed and subsequently reviewed due to an analysis step being disabled with the monitoring service application software.

Source facts

Action
On 12/18/2024, the firm sent an Urgent Medical Device Correction Letter via certified mail to customers informing them that it has been identified a software configuration issue that resulted in a percentage of ECG events received into Monitoring Service Application (MSA) were not properly routed due to the Secondary Analysis being disabled and subsequently there ECG events were not reviewed by a Cardiology Technician for inclusion in reporting. Importantly, some of the ECG events received into MSA during this period (July 2022-July 2024) met notification criteria for escalation to the ordering practitioners but were not escalated. This may have resulted in missing information in the reports or notifications you received, which could have impacted your clinical decision-making. Philips is offering to have the prescription reprocessed to include the ECG events not initially routed to a Cardiology Technician. An Updated Communication to be sent will include Actions to be taken by customers: . Review the list of Frequently Asked Questions relating to this issue. . Log onto the Prescriber Response Site at https://prs.gobio.com with the Location Code provided in the header of this letter. . Read and acknowledge the receipt of this Urgent Medical Device Correction Letter on the website . Review the list of your patients and select whether and on which patients you wish reprocessing to be performed options available include reprocessing All, None, or Selected specific patients on the site. For questions/assistance-contact Prescriber Response Line 888-521-1684.
Initiated
2024-12-18
Root cause
Use error
Distribution
US Nationwide distribution including PR.
Recall number
Z-0708-2025
Recall status
Open, Classified
CDRH record ID
211327
FDA classified
2024-12-23
Recall event number
95837

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗