Open original FDA record ↗Recall record
- FDA identifier
- Z-0709-2014
- Company in FDA record
- King Systems Corp.
- FDA classification date
- Jan 10, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
KLTD214 airways found inside KLTSD414 packages. The KLTSD414 airway has a lumen for use of a gastric suction tube; the KLTD214 airway does not include a lumen for a gastric suction tube.
Source facts
- Action
- December 13, 2013 Urgent Recall Notices were sent to all Direct Accounts identifying the recalled item, risks associated with the incorrectly labeled product and asking customers to stop use, quarantine and destroy any recalled product on hand. Customers were asked to froward the notifcation if the product was further distributed and FAX the completed recall acknowledgement form to 317-776-5175 or email wnethercutt@kingsystems.com. Call the customer service representative (800.262.8673) to arrange for credit of your affected product(s).
- Initiated
- 0013-12-13
- Root cause
- Labeling Change Control
- Terminated
- 2014-05-15
- Distribution
- Distributed in the states of CA, NY
- Recall number
- Z-0709-2014
- Recall status
- Terminated
- CDRH record ID
- 124329
- FDA classified
- 2014-01-10
- Recall event number
- 66869