Z-0709-2025: mint Lesion, Software Versions: 3.9.0 through 3.9.5.

Recall record Z-0709-2025 for Mint Medical GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0709-2025
Company in FDA record
Mint Medical GmbH
FDA classification date
Dec 20, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some software versions have a malfunction where they may show incorrect orientation labels for a specific subset of DICOM images.

Source facts

Action
On 11/13/2024 and 11/15/2024 "URGENT FIELD SAFETY NOTICE" letters were emailed to customers. US customers on 11/13/2024 and OUS customers on 11/15/2024. Actions to be taken by the user: Please read this information carefully and assess whether you are using an affected product version and if DICOM series with differing patient orientation may be used in your organization (Note: mint Lesion" shows a notice above the images when multi-frame images are displayed (see Figures 1 and 2). If that is the case, the malfunction may occur in your system. Please be aware that displayed orientation markers may indicate a wrong patient orientation. Verify the patient orientation based on anatomical landmarks that you can see in the images, when you rely on patient orientation for reporting findings (e.g., "anomaly in left kidney"). If possible, do not assess DICOM data that may lead to the malfunction with mint Lesion" . If you suspect that your organization uses DICOM files with the described parameters and that they have already been read using mint Lesion" , you can contact Mint Medical support (support@mint-medical.com) to request analysis of your mint Lesion" instance. If you believe that this failure could have occurred in past use of mint Lesion", please review the potentially affected radiological reports in your reporting application and take the necessary steps to correct them. Please inform Mint Medical if this may have led to any patient harm in your organization. Actions being taken by the manufacturer The error will be corrected with a software update. Mint Medical Support will contact you when the update is available to schedule the installation of the update on your system. If you have any questions, email: incidents@mint-medical.com or call (+49) 6221 32 18018.
Initiated
2024-11-15
Root cause
Software design
Distribution
Worldwide distribution: US (nationwide): AR, CA, FL, GA, MI, MN, NY, OH, PA, TN, TX, WI; and OUS (International) to countries of: Austria, Belgium, France, Germany, Italy, Netherlands, Spain, Switzerland, United Kingdom.
Recall number
Z-0709-2025
Recall status
Open, Classified
CDRH record ID
211435
FDA classified
2024-12-20
Recall event number
95880

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗