Z-0710-2022: Alinity m Integrated Reaction Units (IRU)

Recall record Z-0710-2022 for Abbott Molecular, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0710-2022
Company in FDA record
Abbott Molecular, Inc.
FDA record creation date
Mar 2, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is potential for the Reaction Vessels (RVs) detaching from the RV cap during transport by the pipettor robot.

Source facts

Action
On February 11, 2022, Abbott Molecular issued a Product Recall letter to all customers who received Integrated Reaction Unit (IRU) List Number 09N26-010 Lot Numbers 766233, 770433, 771200, 771201, and 773642 used with the Alinity m System. Consignees must complete and return the included customer reply form. Customers are instructed to follow the Necessary Actions outlined in the Customer Letter: Discontinue use of and discard all remaining inventory of IRU Lots 766233, 770433, 771200, 771201, and 773642. To coordinate replacement, contact your local Molecular Diagnostics at Abbott representative. If RVs are found inside the instrument with no cap and there are associated 1998 errors, please contact an Abbott representative to inspect the instrument. Results generated between the onset of the associated 1998 errors and the identification of the RV with no cap inside the instrument should be considered presumptive. Retesting of samples that yielded positive results during this timeframe should be considered.
Initiated
2022-02-11
Root cause
Under Investigation by firm
Terminated
2023-04-14
Distribution
US Nationwide distribution in the states of AZ, CA, CO, CT, DC, FL, GA, IL, KY, LA, MA, MI, MO, NC, NJ, NY, OK, RI, SD, TX, WA, and WI.
Recall number
Z-0710-2022
Recall status
Terminated
CDRH record ID
192039
FDA record created
2022-03-02
Recall event number
89603

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗