Z-0710-2024: Product REF 4050-100N, DRAIN, EXPRESS BRU W/AC, Blood Recovery Unit

Recall record Z-0710-2024 for Maquet Cardiovascular, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0710-2024
Company in FDA record
Maquet Cardiovascular, LLC
FDA classification date
Jan 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Express chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.

Source facts

Action
On December 13, 2023, the firm notified customers through "URGENT MEDICAL DEVICE CORRECTION" letters. Customers were instructed to examine their inventory to determine if they had any affected lots of the Atrium Express Dry Suction Dry Seal Chest Drains. The correction only affects the sterile water syringe; do not use this syringe. If customers do not need visualization of active pneumothorax, they may keep the affected Express chest drain and set up without water. If customers wish to use the Express chest drain as intended, they may replace the affected sterile water syringe with a new syringe filled with sterile water using aseptic technique. Customers also have the option of returning affected Express chest drains to the firm for credit. If you have any questions, please contact your Getinge representative or call the Getinge Customer Support at (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m.(Eastern Time Zone).
Initiated
2023-12-13
Root cause
Material/Component Contamination
Distribution
Worldwide distribution: US (nationwide) and OUS (Foreign) to countries of: Australia, Brazil, Canada, Luxembourg, Netherlands, Norway, Spain and UK
Recall number
Z-0710-2024
Recall status
Open, Classified
CDRH record ID
204705
FDA classified
2024-01-12
Recall event number
93548

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗