Open original FDA record ↗Recall record
- FDA identifier
- Z-0712-2024
- Company in FDA record
- Ortho-Clinical Diagnostics, Inc.
- FDA classification date
- Jan 11, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ortho-Clinical Diagnostics, Inc. is recalling QMS Tacrolimus Calibrators, Lot 74859619 expiry date: May 11, 2024, due to improper storage that is expected to degrade the quality of the calibrator and render it unsuitable for use. Use of the affected lot may results in erroneous calibration results, causing delay in testing.
Source facts
- Action
- On 21 November 2023, all consignees who were shipped the VITROS QMS Tacrolimus Calibrators Lot 74859619 expiry date: 11-May-2024 from QuidelOrtho distribution center were notified of the issue and instructed to: -Discontinue using, render unusable, and discard your inventory of QMS Tacrolimus Calibrators, Lot 74859619, which was shipped to the laboratory from QuidelOrtho. -Post the notification by the laboratory s inventory storage area until a replacement lot of QMS Tacrolimus Calibrator has been received. -Forward the notification if the affected product was distributed outside their facility. If you have questions about this recall, please contact QuidelOrtho's Global Services Organization (formerly Ortho Care) at 1-800-421-3311.
- Initiated
- 2023-11-21
- Root cause
- Storage
- Distribution
- Domestic distribution to IL, IN, FL, NJ, NM, TX. International distribution to Chile.
- Recall number
- Z-0712-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204977
- FDA classified
- 2024-01-11
- Recall event number
- 93621