When SonoCalc IMT 3.0 software is used with the SonoSite 180Plus ultrasound system, the software miscalculates the intima-media-thickness, a measurement of risk for cardiovascular or cerebrovascular events. Underestimates percentage of stenosis, false negative result.
Source facts
Action
On October 26, 2005, the company sent their customers a User Guide Addendum that told customers how to verify that the image calibration data was correct.
Initiated
2005-10-26
Root cause
Other
Terminated
2006-04-10
Distribution
Nationwide, Kuwait, Mexico, South Africa, Chile, Spain, Japan, Puerto Rico, Australia.
A list of 155 cases in which the software was sent out lists medical facilities, physicans, and SonoSite Distributors.
Recall number
Z-0713-06
Recall status
Terminated
CDRH record ID
44741
FDA classified
2006-04-06
Recall event number
34808