Open original FDA record ↗Recall record
- FDA identifier
- Z-0713-2014
- Company in FDA record
- Radiometer America Inc
- FDA classification date
- Jan 10, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.
Source facts
- Action
- Radiometer sent a Product Notification letter dated June 18, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory and remove any affected product, discard all affected product in their possession, and complete page 2 of the letter and return to their Radiometer representative. Radiometer will replace any quantity of affected product they have at their location. Customers with questions were instructed to contact Radiometer America Technicial Support at 1-800-736-0600, opt 4. For questions regarding this recall call 440-871-8900, ext 209.
- Initiated
- 2013-06-10
- Root cause
- Mixed-up of materials/components
- Terminated
- 2014-06-19
- Distribution
- Worldwide Distribution - US (nationwide) including states of: CO, CT, GA, MA, MO & MS and country of: Canada.
- Recall number
- Z-0713-2014
- Recall status
- Terminated
- CDRH record ID
- 124601
- FDA classified
- 2014-01-10
- Recall event number
- 67044