- FDA identifier
- Z-0714-2011
- Company in FDA record
- Accuray Inc
- FDA classification date
- Dec 16, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
If users modify the dose calculation box to cover just the contoured anatomy and not the entire CT volume, dosage may not be properly calculated.
Source facts
- Action
- Accuray Inc., sent an "URGENT DEVICE CORRECTION" letter dated October 21, 2010, to all customers. The letter described the product, problem and actions to be taken by the customers. The customers were instructed to increase the size of the dose box to include CT volume, not just the contoured anatomy; recalculate dose at high resolution, and examine the treatment plan to ensure that applicable organs at risk are accounted for and all portions of the patient receive the desired level of doses. If you have any questions or concerns regarding this issue, please do not hesitate to contact Accuray Customer Support at +1-877-668-8667 (USA), +1-408-716-4700 (non-USA) or email: customersupport@accuray.com.
- Initiated
- 2010-10-28
- Root cause
- Other
- Terminated
- 2011-04-01
- Distribution
- USA and countries of: Athens, Belgium, Canada, China, Czech Republic, France, Germany, India, Italy, Japan, Korea, Malaysia, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, Thailand, The Netherland, Turkey, Ukraine, United Kingdom, Vietnam, Venezuela.
- Recall number
- Z-0714-2011
- Recall status
- Terminated
- CDRH record ID
- 95637
- FDA classified
- 2010-12-16
- Recall event number
- 57236