Skip to content

Public FDA record

Z-0714-2013: Medisystems 15 Gauge x 3/5 inch Buttonhole Needle Set Product Code: BH-2035PE Product Usage: Medisystems

Recall record Z-0714-2013 for Medisystems a NX Stage Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0714-2013
Company in FDA record
Medisystems a NX Stage Company
FDA classification date
Jan 24, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A total of 5 (ea) single needle pouches are mislabeled as 15 Ga x1" instead of 15 Ga x 3/5".

Source facts

Action
Medisystems sent a notification letter dated December 20, 2012 to affected customers. The letter identified the affected products, problem and actions to be taken. Customers were instructed to immediately open and inspect inventory for the affected products. Results of the inspection are to be provided to Medisystems and the product is to be returned to Medisystems. For questions call (978) 332-5973.
Initiated
2012-12-20
Root cause
Labeling mix-ups
Terminated
2013-10-10
Distribution
US Nationwide Distribution in the states of NY, PA and TN.
Recall number
Z-0714-2013
Recall status
Terminated
CDRH record ID
115114
FDA classified
2013-01-24
Recall event number
63959

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗