After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.
Source facts
Action
An Urgent Medical Device Recall Letter, dated December 2013, was sent to Users. Once available Drager will update components of the IACS Monitoring Solution System.
Initiated
2013-12-10
Root cause
Software design
Terminated
2015-05-20
Distribution
US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.
Recall number
Z-0714-2014
Recall status
Terminated
CDRH record ID
124627
FDA classified
2014-01-10
Recall event number
67056